RESTING ECG
Microcomputer-controlled resting ECG.
Acquisition: 12 simultaneous leads, Standard or Cabrera
Pre-acquisition: Provides 10 seconds of instantaneous ECG acquisition and printing in 3, 6 & 12 channel formats.
Signal input: Type CF according to IEC
Sampling rate: 1000 Hz
Dynamic range: �10mV
Superimposed DC voltage: (Polarization voltage) �600mV
Paper Speed: 5/10/25/50 mm/s (direct)
Digital Resolution: 5 �V
Frequency response: 0.05 - 150 Hz
Common mode rejection: > 140 dB
Input impedance: > 100 MOhms
Leakage current: < 10 �A
Pacemaker detection sensitivity: All leads, duration 0.1-2.5 ms, amplitude>5 mV
ECG interpretation: Adult and Pediatric
SPIROMETRY
Predicteds: Crapo, Morris, Knudson, Knudson76, Polgar, others. Extrapolated predicted values.
Spirometry Measured Values: FVC, FEV0.5, FEV1, FEV1/FVC, FEF25-75, PEF25%, FEF50%, FEF75%, FIVC, FIV1, FIV1/FIVC, PIF, FIF50%, SVC, ERV, IRV, TV, MVV, RR, TV. Pre vs. post medication comparison
Measurement Range:
Flow: 0 to +- 14 l/s
Volume: 0 to +1 11 l
Flow Impedence: < 0.2 mbar*s/l at 12 l/s
Spirometry Standards: P8000 Power meets or exceeds ATS, OSHA and NIOSH standards
|
SYSTEM SPECIFICATIONS
Size: 39 x 33 x 10 cm / 15" X 13" X 3.9"
Weight: 5 Kg (with paper) / 8.8 lbs
Power: 100-115V or 220-240V, 50/60Hz
Battery: Built in 12V lead acid battery. ~ 4 hours of use per charge.
Power Consumption: Internal printing: 28VA max
Memory: About 45 resting ECG or spirometry files
LCD Display:
Type: Backlit high resolution LCD
Size: 320 x 240 pixel, 3 1/2 x 5"
ECG: Selected lead group, operation mode, lead check, heart rate, AC filter, muscle filter, ADS, gain, speed, name, ID, 3 ECG waveform
Spirometry: Real time flow/volume loop or time/volume curve, parameters with predicted, actual, and % difference values, interpretation, Up to 3 efforts, pre/post comparison
Monitor output: Standard monitor interface
Internal printer: High resolution digital printer
External printer: Parallel port for connection to plain paper printers